1.Test procedures
Weigh two pieces sample (approximately 0.04g), start the test according to the set testing procedures, and calculate recovery rate.
2.Combustion conditions
Temperature: 1100℃ Oxygen: 0.4L/min
Argon(carrier): 0.2L/min Argon flow for humidification: 0.1L/min
Water vapor: 0.1L/min Absorption solution: 0.05% H2O2
Combustion process
Step | Sample Introduction Time (s) | Distance (mm) | Retention Time (s) |
Step 1 | 17 | 170 | 100 |
Step 1 | 100 | 20 | 100 |
Step 3 | 100 | 15 | 100 |
Step 4 | 150 | 105 | 100 |
3.IC conditions
Analytical Column: IonPac AS19 2*250mm Guard Column: IonPac AG19 2*50mm
Eluent: EGC Ш KOH, 20mM, 0.25mL/min Column temp.: 30℃
Quantitative loop: 400ul Linear range:0.1-4mg/L
Detector: Dionex-ADRS 600
4.Test results

Sample | Cl | Br | S | |||
Measured (mg/kg) | Certified (mg/kg) | Measured (mg/kg) | Certified (mg/kg) | Measured (mg/kg) | Certified (mg/kg) | |
ERM-EC681M | 367 | 380±60 | 1422 | 1430±80 | 643 | 640±100 |
378 | 1454 | 633 | ||||
388 | 1433 | 646 | ||||
383 | 1467 | 659 | ||||
372 | 1456 | 636 | ||||
391 | 1434 | 629 | ||||
Aver. | 379.83 |
| 1444.33 |
| 641.00 |
|
RSD(%) | 2.44 |
| 1.19 |
| 1.69 |
|