Français Русский Español Deutsch Bahasa Indonesia Tiếng Việt 한국어 ภาษาไทย हिन्दी العربية Português 日本語 English 中文
Determination of fluorine in Perfluoroalkyl and Polyfluoroalkyl Substances(PFAS)2026-03-19  view:76

1.Test procedures

Weigh a certain amount of perfluorohexyl sulfonic acid (reference materials), dissolve it in methanol, dilute it to an appropriate concentration, transfer a suitable volume of the solution onto the ignited quartz wool, perform the test according to the set procedure, and calculate the recovery rate.

 

2.Combustion conditions

Temperature: 1100℃                 Oxygen: 0.4L/min             

Argon(carrier): 0.2L/min               Argon flow for humidification: 0.1L/min          

Water vapor: 0.1mL/min          Absorption solution: pure water


Instrument:SDXNS200H

Combustion process

Step

Sample Introduction Time (s)

Distance (mm)

Retention Time (s)

Step 1

17

170

100

Step 1

100

20

100

Step 3

100

15

100

Step 4

150

105

100

 

3.IC conditions

Analytical Column: IonPac AS19 2*250mm   Guard Column: IonPac AG19 2*50mm

Eluent: EGC Ш KOH, 20mM, 0.25mL/min    Column temp.: 30℃

Quantitative loop: 400ul                               Linear range:0.1-1mg/L

Detector: Dionex-ADRS 600

 

4.Test results

 


S/N.

CRM

Brand

Type

Theoretical Conc. (ug)

Measured Conc. (ug)

Recovery

(%)

1

perfluorohexyl

sulfonic acid

TM standard

Low conc.

0.3

0.2938

97.93

2

0.2927

97.57

3

0.2968

98.93

4

0.2899

96.63

5

0.2916

97.20

6

0.2987

99.57

Aver.

0.2939

97.97

RSD(%)

1.12

1.12

1

perfluorohexyl

sulfonic acid

TM standard

High conc.

2

1.9278

96.39

2

1.9871

99.36

3

1.9381

96.91

4

1.9787

98.94

5

1.9191

95.96

6

1.9856

99.28

Aver.

1.9561

97.80

RSD(%)

1.59

1.59

                    

Sales Contact
tel 0086 731 88112150
fax 0086 731 88134650
email sales@sandegroup.com
Service Contact
tel 0086 731 89864000
fax 0086 731 88134650
email service@sandegroup.com
Copyright©HunanSundy Science and Technology Co., Ltd.
湘ICP备10204722号