SDCXS100 Calorific-Halogen Analyzer
Product description
Calorimeter /touch scrren
Calorimeter dimension:542mm*400mm*490mm
Calorimeter weight:54kg
Halogen module/computer
Halogen module dimension:492mm×430mm×780mm
Halogen module weight:50kg
Highlights
1.Fully automatic docking of calorific value and halogen analysis, greatly improving test efficiency.
① Automatic oxygen filling and discharging, automatic pressure detection, automatic water volume determination, automatic water temperature adjustment, automatic oxygen bomb washing.
②Automatic absorption of solube vapors, automatic collection of absorbing solution, automatic quantification, automatic storage, automatic dilution, automatic ion chromatography analysis of halogen and sulfur content.
2.Rapid analysis with stable and accurate results.
①The water temperature in the bucket is automatically adjusted, and the test conditions of each sample are very close, which improves the consistency of the test results and ensures the long-term stability and accuracy of the instrument.
②Ion chromatography method eliminates the problem of mutual interference between different ions in the solution, and has low detection limit, high accuracy and good repeatability.
③The calorific value analysis time <7.5min, halogen and sulfur analysis time <35min. Autosampler automatically cache the absorbing solution in order, and the calorific value test efficiency is not affected by the halogen test efficiency.
3.All-in-one designed bomb calorimeter, integrated with a large touc-screen monitor, high environmental adaptability.
①Optimized calorific value analysis system, all-in-one design with touch-screen monitor, no external cooling water tank, easy operation.
②Dual thermoelectric cooler for cooling and heating, making the temperature control of jacket more precise and improving the environmental adaptability.
4.Stable and safe operation, open database can be connected to LIMS.
①The bomb calorimeter has the self-diagnostic function, and the system structure is clear, which is convenient for daily maintenance. Equipped with a unique tool to easily open and close the oxygen bomb.
②The halogen and sulfur analysis module is equipped with an autoloader with 30 sample positions, the collected solutions can be automatically backed up/cached, support sequential and non-sequential analysis, and can automatically calibrate ion chromatography sensors to reduce manual operations.
③Open database that can be connected to LIMS to automatically collect experimental data.
5.Good adaptability for solid and hazardous waste samples analysis.
①Halogen resistant oxygen bomb solves the problem of high corrosiveness of high halogen samples.
②Provide solutions for the analysis of samples with high volatility, high calorific value and high metal salt content, and solve the problem of easy damage to oxygen bombs, crucibles and other accessories.
③Customized combustion bags with stable calorific value to solve the problem of inconvenient testing of volatile and low calorific value samples.
Working principle diagram
Specification:
Calorific value analysis | |
Sample mass: | 0.05~1g |
Analysis time: | <7.5min |
Temperature resolution: | 0.0001℃ |
RSD: | ≤0.10% |
Heat Capacity Stability: | ≤0.20% within 3 months |
Haloge-sulfur analysis | |
Test method: | Ion chromatography |
Analysis time: | <35min (washing – constant volume – dilution – storage – analysis) |
Detection limit: | F 4mg/kg、Cl 26mg/kg、Br 25mg/kg、S 7mg/kg |
Autoloader: | 30 samples |
Max. Power: | 0.75kW |
Power supply: | 220V±10%,50/60Hz |
Standard | HG/T 6116-2022 ,EN 14582:2016,GB/T 30727-2014,GB/T 213-2008,GB/T 384-1981,SN/T 4762-2017,SN/T 3185-2012,GB/T 37861-2019 |