Determination of fluorine in fluororubber2026-03-19  view:58

1.Test procedures

Weigh approximately 0.008g of the sample, perform the test according to the set testing procedure, and calculate the sample recovery rate.


2.Combustion conditions

Temperature: 1100℃                 Oxygen: 0.4L/min             

Argon(carrier): 0.2L/min              Argon flow for humidification: 0.1L/min          

Water vapor: 0.1mL/min             Absorption solution: pure water


Instrument:SDXNS200H

Combustion process

Step

Sample Introduction Time (s)

Distance (mm)

Retention Time (s)

Step 1

17

180

200

Step 1

100

30

100

Step 3

100

20

100

Step 4

30

80

100

 

3.IC conditions

Analytical Column: IonPac AS19 2*250mm   Guard Column: IonPac AG19 2*50mm

Eluent: EGC Ш KOH, 20mM, 0.25mL/min    Column temp.: 30℃

Quantitative loop: 10ul                   Linear range:10-80mg/L

Detector: Dionex-ADRS 600


4.Test results

Sample

Sample mass(g)

Theoretical Conc.(%)

Measured Conc.(%)

Recovery(%)

FKM

0.0081

65

65.01

100.02

0.0083

65.21

100.32

0.0079

65.29

100.45

0.0081

65.19

100.29

0.0078

64.92

99.88

0.0081

65.01

100.02

Aver.

65.11

100.16

RSD(%)

0.22

0.22