0086-731-88112150

Search

sales@sandegroup.com

0086 731 88112150

 

0086 731 89864000

 

0086 731 88134650

 

sales@sandegroup.com

 

Find us in Direct Industry

Sales Contact

Service Contact

0086 731 88134650

 

service@sandegroup.com

 

 

Copyright©HunanSundy Science and Technology Co., Ltd. 2017    湘ICP备10204722号    www.300.cn Changsha Backstage

Hunan Sundy Science and Technology CO.,LTD

 

No.558 West Tongzipo Road,

 

Yuelu Disctrict,

 

Changsha, Hunan Province, China P.R

SDCH425 Carbon & Hydrogen Analyzer

Application:
SDCH425 can be used to determine the carbon and hydrogen content in coal, coke, food, soil, fertilizer etc in the fields of power plant, coal mine, metallurgy, petrochemical industry, steel factory, environmental industry and so on.
描述一
10 Patents. 7 Invention patents included.
描述二
The only type of elemental analyzers with auto combustion crucible replacing device.
没有此类产品
Product description

Highlights:

High Precision.
①With high-performance infrared light supply, superior optical glass as well as pure gold infrared path and precise thermostatic control, the drift of the infrared cell and TCD is greatly optimized.
Humanization Design, Easy Operation.
①Easy and convenient for gas leakage check, operators only need to click on the software.
②Enlarged combustion crucible can greatly reduce the frequency of crucible changing (Approximate every 350 tests).
Intelligent Control.
①Strong data processing capability by easy to use Windows-based software.
②Capable of connecting with network and balance by RS232 interface.
③With CAN bus communication technology, one PC controls several analyzers.
High automation and test efficiency.
①Autoloader with 34 position, automatic sample introduction and test.
②Combustion crucible can be replaced automatically,
without disassembly or cooling.
③Analysis time/sample≤5 minutes, new samples can be added or replaced during the analysis process.
④Calibration every 3 hours or lower frequency by self developed ultra-low drift infrared cell.
Low operation cost.
①Helium consumption is only 0.7L/min for each test. Nitrogen analysis part can be stopped optionally.
(SDCH425 only)
②Oxygen with purity of 99.5% can meet the test requirement.
Specifications:

Description

SDCH425

Measuring Range

C(0.02%~100%) H(0.02%~50%)

Reproducibility

Cd≤1.00%  Hd≤0.25%

Repeatability

Cad≤0.5%  Had≤0.15%

Max.Sample Quantity

34

Analysis time

≤5min/Sample

Sample Mass

75~105mg (Recommended:100mg)

Analysis Method

CH:Infrared Absorption

Combustion Supporting  Gas

Oxygen Purity:≥99.5%      Pressure:≥1MPa

Driving Gas

Nitrogen / Dry Compressed  Air Pressure:≥1MPa <2.5hour

Preheating Time

<2.5hour

Sample Adding/Replacing Function

Available

Auto Analysis

Yes

Crucible Replacing

Automatic

Furnace Temperature

Combination Furnace,950℃±5℃

Air Tightness Detection

Automatic

Humidity

20%to 85%, non-condensing

Ambient Temperature

15 ℃ to 28℃

Power requirement

220 V(-15%~10%)50/60Hz

Max.Power

3.5kW

Dimensions

680*620*880mm

Weight

164kg

Advance Calculation

Yes

Analysis Report

Yes

Statistics Report

Yes

System Monitor

Yes

Standards AS 1038.6.4, ASTM D5373, ASTM D5291, ISO 16634, ISO 16948, ISO 29541, EN 15407,EN ISO 16948:2015-0, GB/T 30728, GB/T 30733, UNI 15104, UNI CEN/TS 15407

 

Corresponding parameter set not found, please add it in property template of background
暂未实现,敬请期待
暂未实现,敬请期待
Previous article
Next article