SDCHN435 Carbon Hydrogen & Nitrogen Analyzer

Application:
SDCHN 435 can be used to determine the carbon, hydrogen and nitrogen content in coal, coke, food, soil, fertilizer etc in the fields of power plant, coal mine, metallurgy, petrochemical industry, steel factory, environmental industry and so on.
描述一
10 Patents. 7 Invention patents included.
描述二
The only type of elemental analyzers with auto combustion crucible replacing device.
没有此类产品
Product description
Highlights:

High Precision.
①With high-performance infrared light supply, superior optical glass as well as pure gold infrared path and precise thermostatic control, the drift of the infrared cell and TCD is greatly optimized.
Humanization Design, Easy Operation.
①Easy and convenient for gas leakage check, operators only need to click on the software.
②Enlarged combustion crucible can greatly reduce the frequency of crucible changing (Approximate every 350 tests).
Intelligent Control.

①Strong data processing capability by easy to use Windows-based software.
②Capable of connecting with network and balance by RS232 interface.
③With CAN bus communication technology, one PC controls several analyzers.
④Separate reagent tube design,reagent can be replaced conveniently and safely.
Patent No.: ZL200710303472.4

High automation and test efficiency.
①Autoloader with 34 position, automatic sample introduction and test.
②Combustion crucible can be replaced automatically,without disassembly or cooling.
③Analysis time/sample≤5 minutes, new samples can be added or replaced during the analysis process.
④Calibration every 3 hours or lower frequency by self developed ultra-low drift infrared cell.
Low operation cost.
①Helium consumption is only 0.7L/min for each test. Nitrogen analysis part can be stopped optionally.
(SDCHN435 only)
②Oxygen with purity of 99.5% can meet the test requirement.


Specifications:

Description

SDCHN435

Measuring Range

C0.02%~100%) H0.02%~50%) N0.01%~50%

Reproducibility

Cd≤1.00%  Hd≤0.25%  Nd≤0.15%

Repeatability

Cad≤0.5%  Had≤0.15%  Nad≤0.08%

Max.Sample Quantity

34

Analysis time

≤5min/Sample

Sample Mass

75~105mgRecommended100mg

Analysis Method

CHInfrared Absorption NThermal Conduction

Combustion Supporting   Gas

Oxygen Purity≥99.5% Pressure≥1MPa

Carrier Gas

Helium Purity≥99.99% Pressure≥1MPa

Driving Gas

Nitrogen / Dry Compressed Air Pressure≥1MPa

Preheating Time

<2.5hour

Sample Adding/Replacing Function

Available

Auto Analysis

Yes

Crucible Replacing

Automatic

Furnace Temperature

Combination Furnace950±5 Reduction Furnace700±5

Air Tightness Detection

Automatic

Humidity

20%to 85%, non-condensing

Ambient Temperature

15  to 28

Power requirement

220 V-15%~10%50/60Hz

Max.Power

4.5kW

Dimensions

680*630*980mm

Weight

204kg

Advance Calculation

Yes

Analysis Report

Yes

Statistics Report

Yes

System Monitor

Yes

 