0086-731-88112150  sales@sandegroup.com

Home>Elemental Analyzer>SDICS450 Infrared Carbon and Sulfur Analyzer

SDICS450 Infrared Carbon and Sulfur Analyzer

Application:

Sundy SDICS450 Infrared Carbon and Sulfur Analyzer can be used to determine the carbon and sulfur content in coal, coke, ash, biomass fuels and other combustibles in fields of power plants, coal mines, paper plants, metallurgy, petrochemical, steel factory etc.

Sundy Sundy

Product description

Highlights

High automation, high test efficiency.

1、Rapid Analysis time of 110-210 seconds.

2、50 samples auto-loader, high automation, no need attendance in the whole process. Manual sample introduction is also available.

3、Patented technology of combustion gas curtain effectively ensures thorough combustion of sample and no emission of gas. Safe and environmental friendly.

4、Filter wool can be replaced automatically, no damage to combustion tube and furnace.

5、Easy and convenient for checking gas leakage, operators only need to click on the software.

 

Humanization design with stable performance.

1、Tailored boat shaped crucible can be placed at random.

2、Automatic identification of crucibles by locator card, no need coding in advance.

 

High precision.

1、Specially designed high temperature combustion tube with secondary combustion function, makes sure the samples combust thoroughly.

2、Self-developed low drift infrared cell guarantees that only one drift in a working day is enough for accurate test.

 

Intelligent control.

1、Easy-to-use Windows-based software: It only needs to click the mouse after weighing and placing the samples.

2、Easy data handling, real time data can be transmitted by internal network.

3、Capable of connecting with network and balance by RS232 interface.

Specifications:

Description SDICS450
Test Method Infrared Absorption
Measuring Range Carbon 0.1%-100%

Sulfur 0.01%-50% (can be extended to 100%)

Standards ISO 29541,ISO 609,ISO 19579 ASTM D1552, ASTM D3177, ASTM D4239, ASTM D5016, ASTM D5373,GB/T 25214, GB/T 28732, AS 1038.6.3.3, BS 1016
Sulfur Resolution 0.001%
Analysis Time/sample 110s-210s
Max. Quantity of sample 50
Sample Weight 100~500mg (100mg recommended)
Furnace Temp 1250℃ – 1350℃
Gas Requirement Oxygen (purity≥99.5%, 99.9% purity is recommended, no electrolytic oxygen)
Temp. Control Precision ±1℃
Temperature Sensor Thermocouple
Power Requirement AC 220V(-15% ~ 10%)50/60Hz
Max. Power 4.5kW
Max Temperature 1450℃
Sample Introduction Automatic
Operational Control Windows®-based software on external PC
Operation Cost Oxygen consumption lowered by intelligent gas supplying system.Desiccant: Dehydrite which is of higher cost
Dimensions 890*670*650mm
Weight 135kg
Advance Calculation Yes
Analysis Report Yes
Statistics Report Yes
System Monitor Yes

Featured Products