SDAF1600b Ash Fusion Tester

Application:

SDAF1600b Ash Fusion Tester can be used to determine the ash fusibility of coal and coke ash, biomass ash, solid wastes ash. It is widely used in power plant, coal industry, metallurgical industry, petrochemical industry, environmental protection, cement industry, paper manufacturing, geological exploration, and institutes.

Dimensions: 550mm*850mm*843mm

Weight:108kg

Specifications
Features
Application Solution
Video
Spare parts
Download

Description

SDAF1600b

Standards

ISO 540,ASTM D1857/D1857M/E935/935M,
GB/T 219, GB/T 30726, AS 1038-1995,BS 1060-1995,UNI CEN/TS 15370

Sample loader16 sample position

Temperature Ramp Rate

From 4°C until 20 °C per minutes, adjustable

Image Collection

1 frame per 1°C

Image Resolution

1280 x 1024 pixels

Temperature display

 °C / °F

Reducing Gas

CO + CO2 / H2 +  CO2

Standard Calibration

Nickel, Gold, Palladium, and Dextrin Powder

Temperature Control Range

Room temperature to 1600℃

Testing time

≦3 hours/ per batch(9 samples)

Resolution

1℃

Temperature Control Precision

±2℃

Test Atmosphere

Oxidizing atmosphere

Weak reducing atmosphere

Power Source

AC 220 V(-15%~10%),50/60 Hz

Dimension

550*850*843mm

Weight

108kg


Highlights

High automation, high test efficiently

1、Unique automatic sample loader. Operator only need to take it from the loader under room temperature.

2、16 sample positions. Fully automatic, test time less than 3 hours per batch.

3、temperature up to1600℃, applicable to different test atmosphere and samples.

 

Optimized design, low operation cost.

1、Designed for both coal and biofuel test. Except coal test, we also equip sample making mould and crucible for ash fusion test of biofuel.

2、CE certified high definition camera and radiator reduce heat radiation and realize vivid picture and long service life.

3、Single layer design for graphite cup increase utilization efficiency of graphite and bituminous coal. Test atmosphere in furnace is ensured.

4、Anti-overflow slot in ash cone support can store spilled ash cone and prevent ash cone stick with the plate.

5、Furnace adopts heat resistant and heat insulation material. Imported material for combustion tube and sample boat.

6、Heating element by silicon molybdenum rod, resist to high temperature (max. temperature 1800℃), antioxidation and low aging.

7、Defogging design minimizes contamination of the observation quartz plate.

8、Simplified structure of the high-temperature tube reduces cracking and extends service life.


Real time monitoring, precise and accurate result.

1、The whole process of the experiment is monitored and recorded in real time, deformationtemperaturecan be automatically determined by the image processing algorithm of the software.

2、Manual observing and camera recording at same port. Good air tightness and low gas consumption.

3、Gas leakage detection and alarm function make the instrument safe and reliable.

 

Smart control, easy to operate.

1、Friendly interface, easy operation, real time display of image and data; All data can be stored, printed and uploaded.

2、Windows-based operation system for automated testing.

 

Inquiry
Please feel free to submit your inquiry information to us. We will contact with you as soon as possible.
Verification code
Sales Contact
tel 0086 731 88112150
fax 0086 731 88134650
email sales@sandegroup.com
Service Contact
tel 0086 731 89864000
fax 0086 731 88134650
email service@sandegroup.com
Copyright©HunanSundy Science and Technology Co., Ltd.
湘ICP备10204722号