Highlights

High precision.

①Unique designed high temperature combustion tube with secondary combustion function, make sure the samples combust thoroughly.

②Adopted high sensitivity and low drift infrared cell,the measurement result is stable and reliable.

③Top quality ultra low drift infrared cell to ensure stability, precision and accuracy.

④Unique gas tightness.

Intelligent control.

①Easy-to-use Windows-based software,It only needs to click the mouse after weighing and placing the samples.

②Easy data handling, real time data can be transmitted by internal network.

③With Ethernet bus communication technology, one PC controls several analyzers.

④Capable of connecting with network and balance by RS232 interface.

Specifications:

DescriptionSDIS401
Test MethodInfrared Absorption
Sulfur Determination Range0.01% – 50%(can be extended to 100%)
StandardsISO 29541,ISO 609,ISO 19579 ASTM D1552, ASTM D3177, ASTM D4239, ASTM D5016, ASTM D5373,GB/T 25214, GB/T 28732, AS 1038.6.3.3, BS 1016
Sulfur Resolution0.001%
Analysis Time/sample< 2 min
Preheating Time≤120min
Sample Loading1
Sample Weight80 – 500mg(100mg Recommended)
Furnace Temp1250℃ – 1350℃
Heating ElementsSilicon Carbide
Temp. Control Precision±1℃
Temperature SensorThermocouple
Power RequirementAC 220V(-15% ~ 10%)50/60Hz
Max. Power4kW
Gas Requirement99.5% purity oxygen≥1MPa
Gas pressureOxygen(0.28±0.01)MPa  Total Air Flow≥1MPa
Sample Introduction and DiscardingManual
Gas Flow CheckingRotor flow meter adopted Gas flow by visual check
Chemical ReagentsDehydrite
Humidity≤85%
Ambient Temperature(15~30)℃
Operational ControlWindows®-based software on external PC
Operation CostOxygen consumption lowered by intelligent gas supplying system.Desiccant: Dehydrite which is of higher cost
Advance CalculationYes
Analysis ReportYes
Statistics ReportYes
System MonitorYes