Sulfur Determination in the Analysis Sample of Biomass2026-03-06  view:135
Sample Preparation
Crush & Grind the sample to size ≤ 1mm, then insert the sample into furnace under 815℃ to burn completely ,then grind the ash to the size ≤ 0.1mm

Instrument Settings
Combustion chamber temp. :1350℃ 
Constant Oven: 48 ±1℃
Oxygen: Purity 99.5%
Pressure: (0.25 ± 0.01) MPa
Oxygen Flow (4.0 ± 0.2) L/min 
Air Extracting Pump: 2.5L/min 
Nominal Sample Weight: 100mg 
Analysis Time: 120s


Instrument:SDIS401


Procedure
1. Prepare instrument as outlined in the operator’s instruction manual (Routine check for temp. stability, gas leak test, etc.)
2. Purge the gas system for 10mins
3. (1-2) waste samples to condition the instrument
4. Run analysis for 1-2 standards sample for verifying the accuracy
5. Test 10-15 samples
6. 1-2 samples of calibration substance for verification (do drift correction if necessary)
7. Repetition of point 5 to point 6 until all samples are processed

Note
As sulfur content in biomass is very low, so need to calibrate the instrument with standard CRM which has extreme low sulfur percentage.

Typical Results

Sample No.

IR Method

Eschka Method

St,d(%)

St,d(%)

1

0.08

0.07

2

0.14

0.14

3

0.05

0.04

4

0.22

0.23